I. Purification SystemThe working principle of a glove box.
The box body and the gas purification system form a sealed working environment. Inert gas or nitrogen is introduced into the box body, and active substances are circulated and removed, ensuring the system ultimately maintains a highly clean and high-purity inert gas environment (both water and oxygen levels can be achieved).1ppmThe following devices).
Mainly forO2 , H2Oof removal.
II. Purification SystemApplication fields of glove boxes
1.Applied in ultra-pure working environments that require the absence of water, oxygen, and dust.
2.Research, development, and production of batteries and battery materials (including lithium-ion batteries, power batteries, solar cells, lithium iron phosphate batteries, etc.).
3.Research and manufacturing of specialized lighting.HDLamps, metal halide lamps, ceramic metal halide lamps, etc.;
4.Welding: Resistance WeldingTIGWelding, laser welding, plasma welding, brazing, etc.;
5.OLEDR&D and production;
6.Research and development, and production of pharmaceutical products;
7.Research and development, as well as production, of supercapacitors;
8.Fine chemicals, nuclear industry, etc.;
9.Research and development, as well as production, of new energy materials.
III.Technical Indicators
| Under standard conditions, the water content inside the tank (Note: Standard conditions refer to one atmosphere of pressure and a constant temperature of 20°C) | ≤1ppm |
| Under standard conditions, the oxygen content inside the box is (Note: Standard conditions refer to one atmosphere of pressure and a constant temperature of 20°C) | ≤1ppm |
The transition chamber can withstand a vacuum level value. | <100pa |
Leakage rate | ≤ 0.05vol%/h |
Leakage rate≤0.05Vol%/ h;
Leak rate conversion:
Assume the volume of the box is1m³, leakage rate of0.05Vol%/h。
Convert to volume:0.05Vol%/ h×1m³=0.0005m³=0.5L, meaning: the leakage rate per hour is 0.5L.
Calculate the hourly oxygen leakage rate, based on the oxygen content in the air being approximately21% calculation:
0.5L×0.21=0.105L; 0.105L÷1000L/m³=0.000105m³。
0.000105×1000000=105ppm。
Therefore, under shutdown conditions, when no air is supplied to the enclosure and the enclosure pressure is maintained at a slight positive pressure:When the oxygen content is not higher than 105 ppm after 1 hour, the leakage rate is considered to be ≤0.05Vol%/h。
IV.The functions implemented by the system
◎ Manual/automatic control of cabinet pressure;
◎ Purification status of the manual/automatic control system;
◎Automatic control of the reduction process in the gas purification system;
◎ Operation errors and warning prompts of the program control system;
◎ System control parameter settings;
◎ Check system history records;
◎System actuator operating condition detection.
◎The touchscreen instantly displays the chamber pressure value, water content value, and oxygen content value.
◎ Reduction and Regeneration: The purification system uses reduction-regenerable purification materials internally, with the reduction process automatically controlled by the system. During reduction and regeneration, it is necessary to supply power and provide a mixed gas of hydrogen (5-10%) and inert gas.
5.Parameters related to the glove box
(I) Glove box configuration and technical parameters:
Model/Model number | Box dimensionsmm (Long)× width × height) | Overall dimensions of the equipmentmm (Long)× Width × Height) | Notes |
GBP800S-2 | ≈1200×800×930 | ≈1915×830×1830 | Monolithic, Material:SUS304 stainless steel, single-sided single station. The purification system is integrated intoOn a single control box。 |
(2) Basic Configuration:
Configuration and Quantity | Technical Parameters | Note |
Standard transition cabin1 piece | DN391×600mm(thickness)3mm) | Located in the box Left Side, Material:SUS304 stainless steel; aluminum alloy door cover, Contains a movable drawer; equipped with a door-opening assist mechanism. |
Tool Transfer Chamber1 piece | DN157×300mm(thickness)3mm) | Located in the box Left Side, Material:SUS304 stainless steel |
Shelf1 piece | Two layers | Back-mounted type |
Box body frame1 piece | High900mm | There is a carbon steel swivel caster for easy movement and fine adjustment of the glove box. |
Transparent visible panel | Thickness:16mm | Using double-layer laminated tempered glass. |
Gloves2Only. | Thickness:0.4mm, cuff diameter 200mm | Butyl synthetic rubberBrand: United StatesNORTH |
Power socket1 | 220V, 10A, 8 outlets | Inside the box |
Lighting system | Brand:Philips | Configure the fluorescent lamp and ballast. |
Spare interface4Here is the English translation of the Chinese word "套": **Set / Cover / Case / Sleeve / Suit** (Note: "套" has multiple meanings depending on context, such as a set of items, a cover or case for an object, or a sleeve. The specific translation would depend on the full sentence.) | KF40 | Material:SUS304 stainless steel; equipped with a quick-connect fitting on the right side of the box for easier liquid or gas entry and exit. |
Cooling water connector | 2 | Φ8 orΦOn a 6mm enclosure |
Vacuum pump | 4L/s | German Leybold Vacuum Pump |
HEPA filter 2 pieces | Filtration precision0.3μm | Installed inside the cabinet to filter dust. |
Vacuum gauge2 | A mechanical dial display | Measure the vacuum level of the transition chamber and the tool transition chamber. |
Pressure gauge1 | The mechanical dial displays | Measure the air pressure inside the box. |
Foot controller1 piece | Double-pedal type | Control the gas replenishment and evacuation of the enclosure. |
Control system | Color touch screen:7 inches | Touch screen,PLC Brand: SIEMENS; Touchscreen mounted on the enclosure, freely rotatable. The Chinese and English operation interfaces can be freely switched. |
Pressure sensor1 | -2500~2500Pa (relative pressure) | TouchscreenDisplay Brand:SETRA |
Water analyzer1 | 0~1000ppm | TouchscreenDisplay, Accuracy:0.1ppm, Brand:MICHELL |
oxygen transmitter1 | 0~1000ppm | TouchscreenDisplay, Accuracy:0.1ppm, ZrO2 Oxygen Sensor Brand: USA All Zirconia Material |
Gas Purification System | Single purification column | Has water and oxygen removal functions; integrated inOn a single control box; Water Removal Materials: United StatesUOP, deoxygenation material:BASF。 |
Circulating Fan | High-frequency fan | Cycle, pumping speed:0-90m³/h Variable Frequency Drive |
| Purchasing Unit Region | Purchasing Unit Name | Purchase Time | Number of units purchased | Applicable Fields |
| Shanghai | Pangang Group Research Institute | 2019-09-15 | 1 | Energy |