Post Demand

Purification System Glove Box

价  格 
99000.00
品牌:
创未来
型号:
GBP 800S-2
Weight (kg):
685KG
Specifications and Dimensions (Length * Width * Height):
1200*800*930
湖南创未来机电设备制造有限公司
店主:邓经理
地址:湖南省岳阳市平江县伍市镇华文路伍市工业园
  • 产品介绍

I. Purification SystemThe working principle of a glove box.

The box body and the gas purification system form a sealed working environment. Inert gas or nitrogen is introduced into the box body, and active substances are circulated and removed, ensuring the system ultimately maintains a highly clean and high-purity inert gas environment (both water and oxygen levels can be achieved).1ppmThe following devices).

Mainly forO2 , H2Oof removal.

II. Purification SystemApplication fields of glove boxes

1.Applied in ultra-pure working environments that require the absence of water, oxygen, and dust.

2.Research, development, and production of batteries and battery materials (including lithium-ion batteries, power batteries, solar cells, lithium iron phosphate batteries, etc.).

3.Research and manufacturing of specialized lighting.HDLamps, metal halide lamps, ceramic metal halide lamps, etc.;

4.Welding: Resistance WeldingTIGWelding, laser welding, plasma welding, brazing, etc.;

5.OLEDR&D and production;

6.Research and development, and production of pharmaceutical products;

7.Research and development, as well as production, of supercapacitors;

8.Fine chemicals, nuclear industry, etc.;

9.Research and development, as well as production, of new energy materials.

III.Technical Indicators

Under standard conditions, the water content inside the tank

(Note: Standard conditions refer to one atmosphere of pressure and a constant temperature of 20°C)

1ppm

Under standard conditions, the oxygen content inside the box is

(Note: Standard conditions refer to one atmosphere of pressure and a constant temperature of 20°C)

1ppm

The transition chamber can withstand a vacuum level value.

100pa

Leakage rate

0.05vol%/h

Leakage rate0.05Vol%/ h

Leak rate conversion:

Assume the volume of the box is1m³, leakage rate of0.05Vol%/h

Convert to volume:0.05Vol%/ h×1m³=0.0005m³=0.5L, meaning: the leakage rate per hour is 0.5L.

Calculate the hourly oxygen leakage rate, based on the oxygen content in the air being approximately21% calculation:

0.5L×0.21=0.105L; 0.105L÷1000L/m³=0.000105m³。

0.000105×1000000=105ppm。

Therefore, under shutdown conditions, when no air is supplied to the enclosure and the enclosure pressure is maintained at a slight positive pressure:When the oxygen content is not higher than 105 ppm after 1 hour, the leakage rate is considered to be ≤0.05Vol%/h

IV.The functions implemented by the system

◎ Manual/automatic control of cabinet pressure;

◎ Purification status of the manual/automatic control system;

◎Automatic control of the reduction process in the gas purification system;

◎ Operation errors and warning prompts of the program control system;

◎ System control parameter settings;

◎ Check system history records;

◎System actuator operating condition detection.

◎The touchscreen instantly displays the chamber pressure value, water content value, and oxygen content value.

◎ Reduction and Regeneration: The purification system uses reduction-regenerable purification materials internally, with the reduction process automatically controlled by the system. During reduction and regeneration, it is necessary to supply power and provide a mixed gas of hydrogen (5-10%) and inert gas.

5.Parameters related to the glove box

(I) Glove box configuration and technical parameters:

Model/Model number

Box dimensionsmm

(Long)× width × height)

Overall dimensions of the equipmentmm

(Long)× Width × Height)

Notes

GBP800S-2

≈1200×800×930

1915×830×1830

Monolithic, Material:SUS304 stainless steel, single-sided single station. The purification system is integrated intoOn a single control box

(2) Basic Configuration:

Configuration and Quantity

Technical Parameters

Note

Standard transition cabin1 piece

DN391×600mm(thickness)3mm

Located in the box Left Side, Material:SUS304 stainless steel; aluminum alloy door cover,

Contains a movable drawer; equipped with a door-opening assist mechanism.

Tool Transfer Chamber1 piece

DN157×300mm(thickness)3mm

Located in the box Left Side, Material:SUS304 stainless steel

Shelf1 piece

Two layers

Back-mounted type

Box body frame1 piece

High900mm

There is a carbon steel swivel caster for easy movement and fine adjustment of the glove box.

Transparent visible panel

Thickness:16mm

Using double-layer laminated tempered glass.

Gloves2Only.

Thickness:0.4mm, cuff diameter 200mm

Butyl synthetic rubberBrand: United StatesNORTH

Power socket1

220V, 10A, 8 outlets

Inside the box

Lighting system

Brand:Philips

Configure the fluorescent lamp and ballast.

Spare interface4Here is the English translation of the Chinese word "套": **Set / Cover / Case / Sleeve / Suit** (Note: "套" has multiple meanings depending on context, such as a set of items, a cover or case for an object, or a sleeve. The specific translation would depend on the full sentence.)

KF40

Material:SUS304 stainless steel; equipped with a quick-connect fitting on the right side of the box for easier liquid or gas entry and exit.

Cooling water connector

2

Φ8 orΦOn a 6mm enclosure

Vacuum pump

4L/s

German Leybold Vacuum Pump

HEPA filter 2 pieces

Filtration precision0.3μm

Installed inside the cabinet to filter dust.

Vacuum gauge2

A mechanical dial display

Measure the vacuum level of the transition chamber and the tool transition chamber.

Pressure gauge1

The mechanical dial displays

Measure the air pressure inside the box.

Foot controller1 piece

Double-pedal type

Control the gas replenishment and evacuation of the enclosure.

Control system

Color touch screen:7 inches

Touch screen,PLC Brand: SIEMENS; Touchscreen mounted on the enclosure, freely rotatable.

The Chinese and English operation interfaces can be freely switched.

Pressure sensor1

-2500~2500Pa (relative pressure)

TouchscreenDisplay Brand:SETRA

Water analyzer1

0~1000ppm

TouchscreenDisplay, Accuracy:0.1ppm, Brand:MICHELL

oxygen transmitter1

0~1000ppm

TouchscreenDisplay, Accuracy:0.1ppm, ZrO2 Oxygen Sensor Brand: USA All Zirconia Material

Gas Purification System

Single purification column

Has water and oxygen removal functions; integrated inOn a single control box

Water Removal Materials: United StatesUOP, deoxygenation material:BASF

Circulating Fan

High-frequency fan

Cycle, pumping speed:0-90m³/h Variable Frequency Drive


Typical users:
Purchasing Unit Region Purchasing Unit Name Purchase Time Number of units purchased Applicable Fields
Shanghai Pangang Group Research Institute 2019-09-15 1 Energy
对比栏关闭